Compare
PG 8583 CD [WW ADP CM]
Laboratory dishwasher with pump for AD water, liquid dispensing, DryPlus & conductivity monitoring.
- Performance/cycle, e.g. 128laboratory bottlesFurther stacking options
Alternatively, 98 pipettes can be reprocessed per cycle with further laboratory glassware. - Safety – Wash pressure, spray arm and conductivity monitoringFor a reliable result
Outstanding preparation results are achieved because the jet pressure and spray arms are monitored continuously and a decline in jet pressure or load blockage is detected immediately In addition, any deviations from the prescribed conductivity of the water, particularly in the final rinse, are detected. - Efficient hot air drying - DryPlusEfficient drying
Hot air drying DryPlus is the optimum solution for hollow instruments or narrow-lumened laboratory glassware. The upstream H14 HEPA filter ensures optimum particle separation from the drying air. The filter is easily accessible via a side unit. - Total security - link to process documentation possibleEnd-to-end documentation
The optional addition of a module for process documentation is provided via a communication slot. The Ethernet module is used for documentation via software. The RS 232 module must be used for documentation within a USB solution or via a printer.
Product benefits
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Clear text display
Text display
Everything at a glance: The display shows everything as clear text in the country-specific language.
AutoClose*Patent: EP 1 733 675 B1
Automatically closing door
A light touch is all it takes: With AutoClose, the door closes by itself.
Large selection of components
Reprocess flexibly
Flexible and comprehensive components for optimum cleaning performance and a variety of reprocessing options.
Can be built under a worktop
For a consistent look
The machine can be easily integrated under a worktop.
Additional integrable dispenser pumps
Dispense practically and easily
For applications that require additional media, additional dispenser pumps can be integrated into the machine.
Subject to technical changes; no liability accepted for the accuracy of the information given! Figures as examples for showing the product benefit.
Product details
PG 8583 CD [WW ADP CM]
EAN 4002515630570 / Article number 62858351GB / Material number 10224290
EAN 4002515630570 / Article number 62858351GB / Material number 10224290
Construction type | Freestanding, can be built-under, wide, without lid Note Factory default without appliance cover: if required, the lid is available separately (for conversion to a freestanding appliance). |
---|---|
Casing | Stainless steel |
Can be built-under | |
AutoClose – automatic door lock | AutoClose* A light touch is all it takes: With AutoClose, the door closes by itself.Patent: EP 1 733 675 B1 |
Buzzer, acoustic signal at end of programme | |
Service-friendly design |
Laboratory flasks per cycle [number] | 128 |
---|---|
Pipettes per cycle [number] | 98 |
Circulation pump, Qmax in l/min | 500 |
---|---|
Maximum final rinse temperature in °C | 93 |
Wash cabinet usable capacity in l | 145 |
Tested operating hours Note Our machines undergo intensive testing under laboratory conditions. The actual operating hours may vary depending on actual usage behaviour, maintenance and care of the machine. | 15000 |
Control type | TouchControl |
---|---|
Programme selection | Touch on Metal Touch on Metal Greater user convenience and easy surface cleaning. |
Direct-access pushbuttons | |
Max. delay start in h | 24 |
Programmability | programmable |
Programmes [number] | 15 |
Free programme positions [number] | 2 |
A0 value controls | |
Automatic programme restart | Programme failure protection Time-saving and safe: The programme failure protection facilitates the continuation of the cleaning and disinfection process after a short interruption, such as a power cut. If the success of the reprocessing process should be at risk, the appliance issues a corresponding warning message. |
Time left display | |
Programme sequence display | |
Selectable display languages |
Electrical connection | 3N AC 400V 50HZ |
---|---|
Electric heater rating in kW | 8.5 |
Total rated load in kW | 9.3 |
Fuse rating in A | 15-16 |
Length of mains connection cable in m Note The length refers to the supply lead on the outside of the appliance. | 1.8 |
Electrical connection | AC 230V 50HZ |
---|---|
Electric heater rating in kW | 5.5 |
Total rated load in kW | 6.3 |
Fuse rating in A | 30-32 |
Electrical connection | AC 230V 50HZ |
---|---|
Electric heater rating in kW | 2.5 |
Total rated load in kW | 3.3 |
Fuse rating in A | 15-16 |
Cold water [number] | 1 |
---|---|
Cold water for steam condenser [number] | 1 |
Hot water [number] Efficient hot water connection Connection to the warm water supply saves a lot of time and energy. | 1 |
VEP – with feed pump for DI water [number] | 1 |
Required flow pressure in kPa Note on ADP variants This value applies to cold, hot and DI water; however, it does not apply to DI water intake on ADP variants because these use a feed pump. | 200-1000 |
Maximum delivery head for drain pump in cm | 100 |
Integrated water softener | |
Maximum water hardness (cold water/hot water) in mmol/l | 10.7 |
Drain pump [DN] | 22 |
Water protection system Waterproof System (WPS) Protection against flooding: In the event of problems, the system stops water intake automatically. | Waterproof system |
Integrated drying unit | |
---|---|
Fan in kW | 0.3 |
Heater element in kW | 2.2 |
Total rated load in kW | 2.5 |
Air throughput in m³/h | 60 |
Temperature setting in 1 °C increments | 50-115 |
Time setting in 1-minute increments | 0-120 |
HEPA filter class | H14 Efficient hot air drying Efficient internal and external drying of the whole instrument using hot air filtered with a H14 HEPA filter. |
HEPA filter filtration efficiency (in accordance with DIN EN 1822) in % | 99.995 |
HEPA filter service life in h | 500 |
External dimensions, net height in mm | 820 |
---|---|
External dimensions, net width in mm | 900 |
External dimensions, net depth in mm | 600 |
External dimensions, gross height in mm Note Including packaging | 1095 |
External dimensions, gross width in mm Note Including packaging | 1200 |
External dimensions, gross depth in mm Note Including packaging | 800 |
Wash cabinet, usable height in mm | 520 |
Wash cabinet, usable width in mm | 530 |
Wash cabinet, depth of upper basket in mm | 474 |
Wash cabinet, depth of lower basket in mm | 520 |
Net weight in kg | 98 |
Gross weight in kg Note Including packaging | 133 |
Emission sound pressure level in the workplace Emissions - sound pressure at workplace The emission sound pressure level depends on the programme and is a value given to the noise created by the machine. Miele Professional appliances have very low noise levels. | <lt/>70 dB(A) re 20 µPa |
---|---|
Heat dissipation into the room in MJ/h Heat output into the room During operation warm air is vented into the surrounding room. The released amount of heat/time depends on the set temperatures and the intensity of use. The stated maximum value can be used to control the temperature of the area or to calculate an increase in room temperature. | 1.4 |
Special 93 °C–10' | |
---|---|
Standard | |
Universal | |
Intensive | |
Inorganic | |
Organic | |
Injector Plus | |
Pipettes | |
Plastics | |
Mini | |
Oil programme |
RS 232 serial interface (optional) | |
---|---|
WiFi via DataDiary (optional) | |
Ethernet interface (optional) |
VDE (electrical safety) | |
---|---|
IP safety class according to EN 60529: IP 21 | |
EN 61010-1 | |
EN 61010-2-040 | |
RoHS guideline 2011/65/EU | |
WRAS | |
Machinery Directive 2006/42/EC |
Additional accessories and consumables
Find additional accessories and consumables to go with this product
Additional accessories for laboratory technology
Find here the accessories that can be bought additionally for this product.
Additional consumables for laboratory technology
Find here the consumables that can be bought additionally for this product.